Field Classification Guide
This guide helps classify instrument data fields into the correct ASM document locations. Use this when mapping raw instrument output to Allotrope Simple Model structure.
Raw lines 1 to 4ASM Document Hierarchy
<technique>-aggregate-document
├── device-system-document # Instrument hardware info
├── data-system-document # Software/conversion info
├── <technique>-document[] # Per-run/sequence data
│ ├── analyst # Who performed the analysis
│ ├── measurement-aggregate-document
│ │ ├── measurement-time
│ │ ├── measurement-document[] # Individual measurements
│ │ │ ├── sample-document
│ │ │ ├── device-control-aggregate-document
│ │ │ └── [measurement fields]
│ │ └── [aggregate-level metadata]
│ ├── processed-data-aggregate-document
│ │ └── processed-data-document[]
│ │ ├── data-processing-document
│ │ └── [processed results]
│ └── calculated-data-aggregate-document
│ └── calculated-data-document[]Raw lines 5 to 27Field Classification Categories
Raw lines 28 to 291. Device/Instrument Information → device-system-document
Hardware and firmware details about the physical instrument.
| Field Type | ASM Field | Examples |
|---|---|---|
| Instrument name | model-number | "Vi-CELL BLU", "NanoDrop One" |
| Serial number | equipment-serial-number | "VCB-12345", "SN001234" |
| Manufacturer | product-manufacturer | "Beckman Coulter", "Thermo Fisher" |
| Firmware version | firmware-version | "v2.1.3" |
| Device ID | device-identifier | "Instrument_01" |
| Brand | brand-name | "Beckman Coulter" |
Rule: If the value describes the physical instrument and doesn't change between runs, it goes in device-system-document.
Raw lines 30 to 46
2. Software/Data System Information → data-system-document
Information about software used for acquisition, analysis, or conversion.
| Field Type | ASM Field | Examples |
|---|---|---|
| Software name | software-name | "Chromeleon", "Gen5" |
| Software version | software-version | "7.3.2" |
| File name | file-name | "experiment_001.xlsx" |
| File path | file-identifier | "/data/runs/2024-01-15/" |
| Database ID | ASM-converter-name | "allotropy v0.1.55" |
Rule: If the value describes software, file metadata, or data provenance, it goes in data-system-document.
Raw lines 47 to 62
3. Sample Information → sample-document
Metadata about the biological/chemical sample being analyzed.
| Field Type | ASM Field | Examples |
|---|---|---|
| Sample ID | sample-identifier | "Sample_A", "LIMS-001234" |
| Sample name | written-name | "CHO Cell Culture Day 5" |
| Sample type/role | sample-role-type | "unknown sample role", "control sample role" |
| Batch ID | batch-identifier | "Batch-2024-001" |
| Description | description | "Protein expression sample" |
| Well position | location-identifier | "A1", "B3" |
Rule: If the value identifies or describes what was measured (not how), it goes in sample-document.
Raw lines 63 to 79
4. Device Control Settings → device-control-aggregate-document
Instrument settings and parameters used during measurement.
| Field Type | ASM Field | Examples |
|---|---|---|
| Injection volume | sample-volume-setting | 10 µL |
| Wavelength | detector-wavelength-setting | 254 nm |
| Temperature | compartment-temperature | 37°C |
| Flow rate | flow-rate | 1.0 mL/min |
| Exposure time | exposure-duration-setting | 500 ms |
| Detector gain | detector-gain-setting | 1.5 |
| Illumination | illumination-setting | 80% |
Rule: If the value is a configurable instrument parameter that affects measurement, it goes in device-control-aggregate-document.
Raw lines 80 to 97
5. Environmental Conditions → device-control-document or technique-specific
Ambient or controlled environmental parameters during measurement.
| Field Type | ASM Field | Examples |
|---|---|---|
| Ambient temperature | ambient-temperature | 22.5°C |
| Humidity | ambient-relative-humidity | 45% |
| Column temperature | compartment-temperature | 30°C |
| Sample temperature | sample-temperature | 4°C |
| Electrophoresis temp | (technique-specific) | 26.4°C |
Rule: Environmental conditions that affect measurement quality go with device control or in technique-specific locations.
Raw lines 98 to 113
6. Raw Measurement Data → measurement-document
Direct instrument readings - the "ground truth" data.
| Field Type | ASM Field | Examples |
|---|---|---|
| Absorbance | absorbance | 0.523 AU |
| Fluorescence | fluorescence | 12500 RFU |
| Cell count | total-cell-count | 2.5e6 cells |
| Peak area | peak-area | 1234.5 mAU·min |
| Retention time | retention-time | 5.67 min |
| Ct value | cycle-threshold-result | 24.5 |
| Concentration (measured) | mass-concentration | 1.5 mg/mL |
Rule: If the value is a direct instrument reading that wasn't computed from other values in this analysis, it goes in measurement-document.
Raw lines 114 to 131
7. Calculated/Derived Data → calculated-data-aggregate-document
Values computed from raw measurements.
| Field Type | ASM Field | Examples |
|---|---|---|
| Viability % | calculated-result | 95.2% |
| Concentration (from std curve) | calculated-result | 125 ng/µL |
| Ratio (260/280) | calculated-result | 1.89 |
| Relative quantity | calculated-result | 2.5x |
| % Recovery | calculated-result | 98.7% |
| CV% | calculated-result | 2.3% |
Calculated data document structure:
{
"calculated-data-name": "viability",
"calculated-result": {"value": 95.2, "unit": "%"},
"calculation-description": "viable cells / total cells * 100"
}Rule: If the value was computed from other measurements in this analysis, it goes in calculated-data-aggregate-document. Include calculation-description when possible.
Raw lines 132 to 157
8. Processed/Analyzed Data → processed-data-aggregate-document
Results from data processing algorithms (peak integration, cell classification, etc.).
| Field Type | ASM Field | Examples |
|---|---|---|
| Peak list | peak-list | Integrated peak results |
| Cell size distribution | cell-diameter-distribution | Histogram data |
| Baseline-corrected data | (in processed-data-document) | Corrected spectra |
| Fitted curve | (in processed-data-document) | Standard curve fit |
**Associated data-processing-document:**
{
"cell-type-processing-method": "trypan blue exclusion",
"cell-density-dilution-factor": {"value": 2, "unit": "(unitless)"},
"minimum-cell-diameter-setting": {"value": 5, "unit": "µm"},
"maximum-cell-diameter-setting": {"value": 50, "unit": "µm"}
}Rule: If the value results from an algorithm or processing method applied to raw data, it goes in processed-data-aggregate-document with its processing parameters in data-processing-document.
Raw lines 158 to 182
9. Timing/Timestamps → Various locations
| Timestamp Type | Location | ASM Field |
|---|---|---|
| Measurement time | measurement-document | measurement-time |
| Run start time | analysis-sequence-document | analysis-sequence-start-time |
| Run end time | analysis-sequence-document | analysis-sequence-end-time |
| Data export time | data-system-document | (custom) |
Rule: Use ISO 8601 format: 2024-01-15T10:30:00Z
Raw lines 183 to 195
10. Analyst/Operator Information → <technique>-document
| Field Type | ASM Field | Examples |
|---|---|---|
| Operator name | analyst | "jsmith" |
| Reviewer | (custom or extension) | "Pending" |
Rule: Analyst goes at the technique-document level, not in individual measurements.
Raw lines 196 to 206
Decision Tree
Is this field about...
THE INSTRUMENT ITSELF?
├── Hardware specs → device-system-document
└── Software/files → data-system-document
THE SAMPLE?
└── Sample ID, name, type, batch → sample-document
INSTRUMENT SETTINGS?
└── Configurable parameters → device-control-aggregate-document
ENVIRONMENTAL CONDITIONS?
└── Temp, humidity, etc. → device-control-document
A DIRECT READING?
└── Raw instrument output → measurement-document
A COMPUTED VALUE?
├── From other measurements → calculated-data-document
└── From processing algorithm → processed-data-document
TIMING?
├── When measured → measurement-document.measurement-time
└── When run started/ended → analysis-sequence-document
WHO DID IT?
└── Operator/analyst → <technique>-document.analystRaw lines 207 to 239Common Instrument-to-ASM Mappings
Note: These mappings are derived from the Benchling allotropy library. For authoritative mappings, consult the parser source code for your specific instrument.Raw lines 240 to 243
Cell Counter (Vi-CELL BLU)
*Source: allotropy/parsers/beckman_vi_cell_blu/vi_cell_blu_structure.py*
| Instrument Field | ASM Field |
|---|---|
| Sample ID | sample_identifier |
| Analysis date/time | measurement_time |
| Analysis by | analyst |
| Viability (%) | viability |
| Viable (x10^6) cells/mL | viable_cell_density |
| Total (x10^6) cells/mL | total_cell_density |
| Cell count | total_cell_count |
| Viable cells | viable_cell_count |
| Average diameter (μm) | average_total_cell_diameter |
| Average viable diameter (μm) | average_live_cell_diameter |
| Average circularity | average_total_cell_circularity |
| Cell type | cell_type_processing_method (data-processing) |
| Dilution | cell_density_dilution_factor (data-processing) |
| Min/Max Diameter | minimum/maximum_cell_diameter_setting (data-processing) |
Spectrophotometer (NanoDrop)
| Instrument Field | ASM Field |
|---|---|
| Sample Name | sample_identifier |
| A260, A280 | absorbance (with wavelength) |
| Concentration | mass_concentration |
| 260/280 ratio | a260_a280_ratio |
| Pathlength | pathlength |
Plate Reader
| Instrument Field | ASM Field |
|---|---|
| Well | location_identifier |
| Sample Type | sample_role_type |
| Absorbance/OD | absorbance |
| Fluorescence | fluorescence |
| Plate ID | container_identifier |
Chromatography (HPLC)
| Instrument Field | ASM Field |
|---|---|
| Sample ID | sample_identifier |
| Injection Volume | injection_volume |
| Retention Time | retention_time |
| Peak Area | peak_area |
| Peak Height | peak_height |
| Column Temp | column_oven_temperature |
| Flow Rate | flow_rate |
Unit Handling
Only use units explicitly present in source data. If a value has no unit specified:
- Use
(unitless)as the unit value - Do NOT infer units based on domain knowledge
Calculated Data Traceability
When creating calculated values, always link them to their source data using data-source-aggregate-document:
{
"calculated-data-name": "DIN",
"calculated-result": {"value": 5.8, "unit": "(unitless)"},
"calculated-data-identifier": "TEST_ID_147",
"data-source-aggregate-document": {
"data-source-document": [{
"data-source-identifier": "TEST_ID_145",
"data-source-feature": "sample"
}]
}
}This declares: "DIN 5.8 was calculated from the sample at TEST_ID_145."
Why this matters:
- Audits: Prove a value came from specific raw data
- Debugging: Trace unexpected results back to their source
- Reprocessing: Know which inputs to re-analyze if algorithms change
Assign unique IDs to:
- Measurements, peaks, regions, and calculated values
- Use a consistent naming pattern (e.g.,
INSTRUMENT_TYPE_TEST_ID_N)
This enables bidirectional traversal: trace from calculated → raw, or raw → all derived values.
Raw lines 299 to 331
Nested Document Structure (Critical)
A common mistake is "flattening" fields directly onto measurement documents when they should be wrapped in nested structures. This breaks schema compliance and loses semantic context.
Raw lines 332 to 335Why Nesting Matters
ASM uses nested documents for semantic grouping:
| Document | Purpose | Contains |
|---|---|---|
sample document | What was measured | Sample ID, locations, plate identifiers |
device control aggregate document | How instrument operated | Settings, parameters, techniques |
custom information document | Vendor-specific fields | Non-standard fields that don't map to ASM |
Sample Document Fields
These fields MUST be inside sample document, not flattened on measurement:
// ❌ WRONG - Fields flattened on measurement
{
"measurement identifier": "TEST_001",
"sample identifier": "Sample_A",
"location identifier": "A1",
"absorbance": {"value": 0.5, "unit": "(unitless)"}
}
// ✅ CORRECT - Fields nested in sample document
{
"measurement identifier": "TEST_001",
"sample document": {
"sample identifier": "Sample_A",
"location identifier": "A1",
"well plate identifier": "96WP001"
},
"absorbance": {"value": 0.5, "unit": "(unitless)"}
}Fields belonging in sample document:
sample identifier- Sample ID/namewritten name- Descriptive sample namebatch identifier- Batch/lot numbersample role type- Standard, blank, control, unknownlocation identifier- Well position (A1, B3, etc.)well plate identifier- Plate barcodedescription- Sample description
Device Control Document Fields
Instrument settings MUST be inside device control aggregate document:
// ❌ WRONG - Device settings flattened
{
"measurement identifier": "TEST_001",
"device identifier": "Pod1",
"technique": "Custom",
"volume": {"value": 26, "unit": "μL"}
}
// ✅ CORRECT - Settings nested in device control
{
"measurement identifier": "TEST_001",
"device control aggregate document": {
"device control document": [{
"device type": "liquid handler",
"device identifier": "Pod1"
}]
},
"aspiration volume": {"value": 26, "unit": "μL"}
}Fields belonging in device control:
device type- Type of devicedevice identifier- Device IDdetector wavelength setting- Wavelength for detectioncompartment temperature- Temperature settingsample volume setting- Volume settingflow rate- Flow rate setting
Custom Information Document
Vendor-specific fields that don't map to standard ASM terms go in custom information document:
"device control document": [{
"device type": "liquid handler",
"custom information document": {
"probe": "2",
"pod": "Pod1",
"source labware name": "Inducer",
"destination labware name": "GRP1"
}
}]Raw lines 414 to 429Liquid Handler: Transfer Pairing
For liquid handlers, a measurement represents a complete transfer (aspirate + dispense), not separate operations:
// ❌ WRONG - Separate records for aspirate and dispense
[
{"measurement identifier": "OP_001", "transfer type": "Aspirate", "volume": {"value": 26, "unit": "μL"}},
{"measurement identifier": "OP_002", "transfer type": "Dispense", "volume": {"value": 26, "unit": "μL"}}
]
// ✅ CORRECT - Single record with source and destination
{
"measurement identifier": "TRANSFER_001",
"sample document": {
"source well location identifier": "1",
"destination well location identifier": "2",
"source well plate identifier": "96WP001",
"destination well plate identifier": "96WP002"
},
"aspiration volume": {"value": 26, "unit": "μL"},
"transfer volume": {"value": 26, "unit": "μL"}
}Pairing logic:
- Match aspirate and dispense operations by probe number
- Create one measurement per matched pair
- Use
source_*fields for aspirate location - Use
destination_*fields for dispense location - Include both
aspiration volumeandtransfer volume
Quick Reference: Nesting Decision
Is this field about...
THE SAMPLE BEING MEASURED?
├── Sample ID, name, batch → sample document
├── Well position → sample document.location identifier
├── Plate barcode → sample document.well plate identifier
└── Source/destination locations → sample document (with prefixes)
INSTRUMENT SETTINGS?
├── Standard settings → device control aggregate document
└── Vendor-specific → custom information document
A MEASUREMENT VALUE?
└── Direct on measurement document (e.g., absorbance, volume)
TRANSFER OPERATION TYPE?
└── DON'T use "transfer type" - pair into single measurement
with source/destination fields insteadRaw lines 462 to 484Validation
Use validate_asm.py to check for nesting issues:
python scripts/validate_asm.py output.json --reference known_good.jsonThe validator checks for:
- Fields incorrectly flattened on measurements
- Missing
sample documentwrapper - Missing
device control aggregate documentwrapper - Missing
custom information documentfor vendor fields - Liquid handler: separate transfer types instead of paired records