Signal processing (pyOpenMS 3.6.0)
Raw lines 1 to 2Decide what the signal represents
Use spec.getType() and acquisition/conversion provenance. Its enum is ms.SpectrumSettings.SpectrumType.PROFILE, CENTROID, or UNKNOWN. isSorted() tests m/z ordering only. getType(True) can estimate type from the signal when metadata is insufficient, but an estimate is not acquisition proof.
process_spectra.py applies smoothing, picking, normalization, S/N filtering, then intensity thresholding to selected MS levels; chromatograms pass through. It rejects centroid spectra for smoothing/picking and requires --assume-profile for independently verified unknown-type spectra. The native tests check a synthetic Gaussian profile, selected-level isolation, metadata/peak-array retention and writeback of filtered spectra.
python scripts/process_spectra.py profile.mzML picked.mzML --ms-level 1 --pick
python scripts/process_spectra.py spectra.mzML scaled.mzML --ms-level 2 --normalize to_one --threshold 0.01These are file-dependent templates. Parameters need representative instrument QC; smoothing can distort narrow peaks. PeakPickerHiRes is for well sampled high-resolution profile peaks. Do not repeatedly centroid already picked data.
API patterns
import pyopenms as ms
import numpy as np
spec = ms.MSSpectrum()
spec.setType(ms.SpectrumSettings.SpectrumType.PROFILE)
mz = np.linspace(499.9, 500.1, 101)
spec.set_peaks((mz, 1000 * np.exp(-0.5 * ((mz - 500) / 0.01) ** 2)))
spec.sortByPosition()
picker = ms.PeakPickerHiRes()
picked = ms.MSSpectrum()
picker.pick(spec, picked)
assert len(picked) == 1
assert abs(picked[0].getMZ() - 500) < 0.001GaussFilter.filter(spec) uses gaussian_width in m/z units for a spectrum; use_ppm_tolerance selects ppm_tolerance instead. Chromatogram widths use seconds. The 3.6 filterSpectrum compatibility alias is broken for these two smoothers; use their native filter(spec) method. SavitzkyGolayFilter uses a point-count frame_length and polynomial_order; validate sampling regularity and window size. PeakPickerIterative has signal_to_noise_, peak_width, and nr_iterations_ parameters; this alternative was signature/parameter checked, not benchmarked on low-resolution instrument data.
Normalization (Normalizer.filterSpectrum) supports to_one and to_TIC. Thresholds afterwards are on normalized intensities; a threshold of 10 after unit-TIC normalization removes all nonnegative peaks. Within-scan normalization usually destroys the abundance information needed for label-free quantification.
For filtering, use spec.select(indices) to keep float/integer/string data arrays synchronized. set_peaks() alone does not filter associated arrays. Spectra read with getSpectrum, indexing or iteration may be copies: explicitly replace the full spectrum list with exp.setSpectra(processed).
Other algorithms and alignment
ThresholdMower, WindowMower, NLargest, and MorphologicalFilter expose parameters through getParameters(); inspect their installed filterSpectrum signatures before use. Their particular noise/baseline assumptions are not validated by a successful call. For SpectraMerger, block-wise merging uses block_method:* options; setting average_gaussian:* while calling mergeSpectraBlockWise does not configure Gaussian averaging.
Deisotoper.deisotopeAndSingleCharge acts on a sorted centroided spectrum; record tolerance units, charge range, isotope count, and whether intensity is summed or masses are converted to singly charged ions. This is not calibration. For feature deconvolution use a FeatureMap and FeatureDeconvolution.compute with output map/groups/edges; there is no potential_charge_states parameter.
For RT alignment, call aligner.setReference(reference_map), aligner.align(sample_map, trafo), then MapAlignmentTransformer().transformRetentionTimes(sample_map, trafo, True). See the feature reference and bundled alignment script. Retention alignment does not correct mass errors.
Calibration: verified signature, illustrative workflow
InternalCalibration.calibrate(exp, reference_masses) is not a valid API. First collect calibrants using fillCalibrants, then fit/apply a model. The following is a template requiring independently established ion m/z values and a representative run; it was not validated on instrument data:
calibration = ms.InternalCalibration()
locks = [ms.InternalCalibration_LockMass(500.0, 1, 1),
ms.InternalCalibration_LockMass(1000.0, 1, 1)]
found, failed = calibration.fillCalibrants(exp, locks, 20.0, False, False, False)
if found < 2:
raise ValueError("Insufficient calibrants; do not fit")
ok = calibration.calibrate(exp, [1], ms.MZTrafoModel.MODELTYPE.LINEAR,
-1.0, False, 5.0, 5.0)
if not ok:
raise ValueError("Calibration residual criteria failed")Lock masses include adducts and require correct MS level and charge. Two points are merely a minimal algebraic requirement for a line; use adequate mass/RT coverage, independent validation ions, outlier diagnostics, isotope checks and before/after ppm residuals. Do not calibrate on unverified accurate-mass database candidates and then use the reduced error as independent identification evidence. Negative rt_chunk selects one global model; positive windows are seconds. R is needed only for requested calibration plot output, not this no-plot template.
Sources: InternalCalibration release source, OpenMS changes, and installed 3.6.0 binding docstrings/defaults.