1# ICH Q2(R2) — Structure and Recommended Data
2
3Research basis: **2026-09-30**, read from the ICH Harmonised Guideline *Validation of Analytical
4Procedures Q2(R2)*, Final Version adopted 1 November 2023, with the error correction dated
530 November 2023. ICH licenses its documents for reuse with acknowledgement, so requirements are
6summarised here directly. Confirm the current text and your region's implementation date at
7<https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130.pdf>.
8

ICH Q2(R2) — Structure and Recommended Data

Research basis: 2026-09-30, read from the ICH Harmonised Guideline Validation of Analytical Procedures Q2(R2), Final Version adopted 1 November 2023, with the error correction dated 30 November 2023. ICH licenses its documents for reuse with acknowledgement, so requirements are summarised here directly. Confirm the current text and your region's implementation date at <https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130.pdf>.

Raw lines 1 to 8
9## Document history that matters
11| Version | Date | Note |
12| --- | --- | --- |
13| Q2A | Oct 1994 | Text |
14| Q2B | Nov 1996 | Methodology |
15| Q2(R1) | Nov 2005 | Q2B merged into the parent guideline |
16| Q2(R2) | 1 Nov 2023 | Complete revision, aligned with the new Q14 |
17| Q2(R2) correction | 30 Nov 2023 | Table 5 reportable-range linearity formulae; Tables 6–11 |
19If a protocol cites "ICH Q2(R1)" or lists characteristics in the R1 order, it is working from the
20superseded structure. The error correction is easy to miss and applies to the dissolution example
21and to Annex 2 Tables 6–11.

Document history that matters

VersionDateNote
Q2AOct 1994Text
Q2BNov 1996Methodology
Q2(R1)Nov 2005Q2B merged into the parent guideline
Q2(R2)1 Nov 2023Complete revision, aligned with the new Q14
Q2(R2) correction30 Nov 2023Table 5 reportable-range linearity formulae; Tables 6–11

If a protocol cites "ICH Q2(R1)" or lists characteristics in the R1 order, it is working from the superseded structure. The error correction is easy to miss and applies to the dissolution example and to Annex 2 Tables 6–11.

Raw lines 9 to 22
23## The restructure
25Q2(R1) presented a flat list. Q2(R2) groups methodology under section 3 by performance
26characteristic:
28```
293.1 Specificity/Selectivity
30 3.1.1 General considerations (absence of interference, orthogonal comparison,
31 technology-inherent justification)
323.2 Range <-- parent characteristic
33 3.2.2 Response
34 3.2.2.1 Linear response <-- what R1 called "linearity"
35 3.2.2.2 Non-linear response
36 3.2.2.3 Multivariate calibration
37 3.2.3 Validation of lower range limits <-- what R1 called LOD and LOQ
383.3 Accuracy and Precision
39 3.3.1 Accuracy
40 3.3.2 Precision (repeatability, intermediate precision, reproducibility)
41 3.3.3 Combined approaches for accuracy and precision <-- new
423.4 Robustness --> largely a development activity, see ICH Q14
43```
45Section 2 carries the general considerations, including two concepts absent from R1: **reportable
46range** (2.3) and **considerations for multivariate analytical procedures** (2.5).

The restructure

Q2(R1) presented a flat list. Q2(R2) groups methodology under section 3 by performance characteristic:

3.1  Specificity/Selectivity
       3.1.1 General considerations (absence of interference, orthogonal comparison,
             technology-inherent justification)
3.2  Range                                    <-- parent characteristic
       3.2.2 Response
               3.2.2.1 Linear response        <-- what R1 called "linearity"
               3.2.2.2 Non-linear response
               3.2.2.3 Multivariate calibration
       3.2.3 Validation of lower range limits <-- what R1 called LOD and LOQ
3.3  Accuracy and Precision
       3.3.1 Accuracy
       3.3.2 Precision (repeatability, intermediate precision, reproducibility)
       3.3.3 Combined approaches for accuracy and precision   <-- new
3.4  Robustness  --> largely a development activity, see ICH Q14

Section 2 carries the general considerations, including two concepts absent from R1: reportable range (2.3) and considerations for multivariate analytical procedures (2.5).

Raw lines 23 to 47
48## Table 1 — which tests for which measured attribute
50Table 1 identifies tests normally conducted by measured quality attribute. It allows justified
51alternatives; the table is not an unconditional legal requirement.
53| Characteristic | Identity | Impurity: quantitative | Impurity: limit test | Assay (content/potency) |
54| --- | --- | --- | --- | --- |
55| Specificity test | yes | yes | yes | yes |
56| Response (calibration model) | no | yes | no | yes |
57| Lower range limit | no | QL† | DL | no |
58| Accuracy test | no | yes | no | yes |
59| Repeatability test | no | yes | no | yes |
60| Intermediate precision test | no | yes‡ | no | yes‡ |
62† In some complex cases DL may also be evaluated.
63‡ Not required independently where reproducibility has been performed and intermediate precision
64can be derived from that dataset.
66Further notes from Table 1: other quantitative measurements follow the impurity scheme when the
67range limit is close to DL/QL, and the assay scheme when it is not. Some characteristics may be
68substituted by technology-inherent justification for physicochemical properties. Lack of specificity
69in one procedure should be compensated by one or more supporting procedures unless justified.

Table 1 — which tests for which measured attribute

Table 1 identifies tests normally conducted by measured quality attribute. It allows justified alternatives; the table is not an unconditional legal requirement.

CharacteristicIdentityImpurity: quantitativeImpurity: limit testAssay (content/potency)
Specificity testyesyesyesyes
Response (calibration model)noyesnoyes
Lower range limitnoQL†DLno
Accuracy testnoyesnoyes
Repeatability testnoyesnoyes
Intermediate precision testnoyes‡noyes‡

† In some complex cases DL may also be evaluated. ‡ Not required independently where reproducibility has been performed and intermediate precision can be derived from that dataset.

Further notes from Table 1: other quantitative measurements follow the impurity scheme when the range limit is close to DL/QL, and the assay scheme when it is not. Some characteristics may be substituted by technology-inherent justification for physicochemical properties. Lack of specificity in one procedure should be compensated by one or more supporting procedures unless justified.

Raw lines 48 to 70
71## Table 2 — reportable range examples
73The reportable range derives from the specification and must include the upper and lower
74specification or reporting limits. Other ranges are acceptable if justified; at low amounts a wider
75upper range may be more practical.
77| Use | Low end | High end |
78| --- | --- | --- |
79| Assay of a product | 80% of declared content, or 80% of the lower specification limit | 120% of declared content, or 120% of the upper specification limit |
80| Potency | lowest specification limit −20% | highest specification limit +20% |
81| Content uniformity | 70% of declared content | 130% of declared content |
82| Dissolution, IR, one point | Q − 45% of the lowest strength specification | 130% of declared content of the highest strength |
83| Dissolution, IR, multi-point | lower limit as justified, or QL | 130% of declared content of the highest strength |
84| Dissolution, modified release | lower limit as justified, or QL | 130% of declared content of the highest strength |
85| Impurity | reporting threshold | 120% of the specification limit |
86| Purity (area %) | 80% of the lower specification limit | upper specification limit, or 100% |
88Where assay and impurity run as a single test with one standard, linearity must be shown both at the
89impurity reporting level and up to 120% of the assay specification limit.
91**Reportable range vs working range.** The reportable range is the interval of *reported results*.
92A working range is what is presented to the instrument, and may differ because of dilution or other
93sample preparation. They can be identical. Mathematical calculation normally links the two.

Table 2 — reportable range examples

The reportable range derives from the specification and must include the upper and lower specification or reporting limits. Other ranges are acceptable if justified; at low amounts a wider upper range may be more practical.

UseLow endHigh end
Assay of a product80% of declared content, or 80% of the lower specification limit120% of declared content, or 120% of the upper specification limit
Potencylowest specification limit −20%highest specification limit +20%
Content uniformity70% of declared content130% of declared content
Dissolution, IR, one pointQ − 45% of the lowest strength specification130% of declared content of the highest strength
Dissolution, IR, multi-pointlower limit as justified, or QL130% of declared content of the highest strength
Dissolution, modified releaselower limit as justified, or QL130% of declared content of the highest strength
Impurityreporting threshold120% of the specification limit
Purity (area %)80% of the lower specification limitupper specification limit, or 100%

Where assay and impurity run as a single test with one standard, linearity must be shown both at the impurity reporting level and up to 120% of the assay specification limit.

Reportable range vs working range. The reportable range is the interval of reported results. A working range is what is presented to the instrument, and may differ because of dilution or other sample preparation. They can be identical. Mathematical calculation normally links the two.

Raw lines 71 to 94
95## Recommended data, by characteristic
97**Specificity (3.1).** Demonstrate absence of relevant interference, or compare against an
98orthogonal procedure, or justify from the technology. For a stability-indicating claim (2.4),
99include samples containing relevant degradation products: spiked with target analytes and known
100interferences, stressed physically and chemically, and aged or stress-stored product samples.
102**Response — linear (3.2.2.1).** Evaluate across the range. **A minimum of five concentrations,
103appropriately distributed, is recommended.** Report the plot, the correlation coefficient or
104coefficient of determination, the y-intercept, the slope, and *an analysis of the deviation of the
105actual data points from the regression line* — for a linear response, assess the impact of any
106non-random pattern in the residual plot. Data may be transformed (for example logarithmically) if
107necessary. Other approaches require justification.
109**Response — non-linear (3.2.2.2).** Some procedures are legitimately non-linear; immunoassays and
110cell-based assays commonly give an S-shaped curve, typically modelled with four- or five-parameter
111logistic functions. For these, **linearity of the concentration–response relationship is not
112required.** Assess the model by non-linear regression, and evaluate whether results are proportional
113to the true values across the range.
115**Response — multivariate (3.2.2.3).** Algorithms may be linear or non-linear. Accuracy depends on
116the distribution of calibration samples across the range and on the reference procedure's error.
117Assess how the residuals change across the calibration range, graphically.
119**Lower range limits (3.2.3).** Four approaches:
121| Approach | DL | QL |
122| --- | --- | --- |
123| Visual evaluation (3.2.3.1) | lowest reliably detected | lowest reliably quantitated |
124| Signal-to-noise (3.2.3.2) | S/N 3:1 generally acceptable | S/N at least 10:1 |
125| SD of response and slope (3.2.3.3) | 3.3σ / S | 10σ / S |
126| Accuracy and precision at the limit (3.2.3.4) | — | validated directly, not estimated |
128σ may come from the SD of blank responses, the residual SD of the regression line, or the SD of
129y-intercepts of regression lines. S is the calibration slope. Signal-to-noise applies only where
130there is baseline noise, and the noise region should sit around where the peak would appear.
132Reporting (3.2.3.5): give the limit **and the approach used**. An estimated limit should then be
133validated by analysing a suitable number of samples at or near it. **For impurity tests the QL must
134be at or below the reporting threshold.** Where the QL is well below the reporting limit — roughly
135ten times lower — the confirmatory validation may be omitted with justification.
137**Accuracy (3.3.1).** Establish across the reportable range under regular test conditions, including
138the sample matrix and the described preparation steps. Three routes: comparison against a reference
139material of known purity, a spiking study into matrix, or comparison against an orthogonal
140procedure. Accuracy can be inferred once precision, response within the range, and specificity are
141established.
143Recommended data (3.3.1.4): an appropriate number of determinations and levels across the reportable
144range — **for example 3 concentrations × 3 replicates of the full procedure.** Report as mean percent
145recovery of a known added amount, or as the difference between the mean and the accepted true value,
146**together with an appropriate 100(1−α)% confidence interval** or justified alternative interval. The
147observed interval should be compatible with the accuracy criterion. For impurities, state whether
148the determination is weight/weight or area percent. For quantitative multivariate procedures use
149RMSEP, compared against an acceptable RMSEC.
151**Precision (3.3.2).** Use authentic homogeneous samples, or artificially prepared ones if
152unavailable.
154- *Repeatability (3.3.2.1)*: **a minimum of 9 determinations covering the reportable range** (for
155 example 3 concentrations × 3 replicates), **or a minimum of 6 determinations at 100% of the test
156 concentration.**
157- *Intermediate precision (3.3.2.2)*: establish the effects of random events — typically different
158 days, environmental conditions, analysts, and equipment. **Studying these effects individually is
159 not necessary**, and design of experiments is encouraged. The extent should be justified from
160 development understanding and risk assessment (ICH Q14).
161- *Reproducibility (3.3.2.3)*: an inter-laboratory trial. **Usually not required for a regulatory
162 submission**, but consider it for pharmacopoeial standardisation or multi-site procedures.
164Recommended data (3.3.2.4): report the standard deviation, the relative standard deviation, and an
165appropriate 100(1−α)% confidence interval.
167**Combined accuracy and precision (3.3.3).** Instead of separate criteria, assess total impact
168against a single combined criterion, using a prediction interval, a tolerance interval, or a
169confidence interval. Report the combined value, describe the approach, and supply the individual
170results as supplemental information where they help justify suitability.
172**Robustness (3.4).** Deliberate variation of procedure parameters, plus stability of sample
173preparations and reagents over the duration of the procedure. Considered during development; may be
174submitted as development data case-by-case or made available on request. See ICH Q14 section 5.

Recommended data, by characteristic

Specificity (3.1). Demonstrate absence of relevant interference, or compare against an orthogonal procedure, or justify from the technology. For a stability-indicating claim (2.4), include samples containing relevant degradation products: spiked with target analytes and known interferences, stressed physically and chemically, and aged or stress-stored product samples.

Response — linear (3.2.2.1). Evaluate across the range. A minimum of five concentrations, appropriately distributed, is recommended. Report the plot, the correlation coefficient or coefficient of determination, the y-intercept, the slope, and an analysis of the deviation of the actual data points from the regression line — for a linear response, assess the impact of any non-random pattern in the residual plot. Data may be transformed (for example logarithmically) if necessary. Other approaches require justification.

Response — non-linear (3.2.2.2). Some procedures are legitimately non-linear; immunoassays and cell-based assays commonly give an S-shaped curve, typically modelled with four- or five-parameter logistic functions. For these, linearity of the concentration–response relationship is not required. Assess the model by non-linear regression, and evaluate whether results are proportional to the true values across the range.

Response — multivariate (3.2.2.3). Algorithms may be linear or non-linear. Accuracy depends on the distribution of calibration samples across the range and on the reference procedure's error. Assess how the residuals change across the calibration range, graphically.

Lower range limits (3.2.3). Four approaches:

ApproachDLQL
Visual evaluation (3.2.3.1)lowest reliably detectedlowest reliably quantitated
Signal-to-noise (3.2.3.2)S/N 3:1 generally acceptableS/N at least 10:1
SD of response and slope (3.2.3.3)3.3σ / S10σ / S
Accuracy and precision at the limit (3.2.3.4)—validated directly, not estimated

σ may come from the SD of blank responses, the residual SD of the regression line, or the SD of y-intercepts of regression lines. S is the calibration slope. Signal-to-noise applies only where there is baseline noise, and the noise region should sit around where the peak would appear.

Reporting (3.2.3.5): give the limit and the approach used. An estimated limit should then be validated by analysing a suitable number of samples at or near it. For impurity tests the QL must be at or below the reporting threshold. Where the QL is well below the reporting limit — roughly ten times lower — the confirmatory validation may be omitted with justification.

Accuracy (3.3.1). Establish across the reportable range under regular test conditions, including the sample matrix and the described preparation steps. Three routes: comparison against a reference material of known purity, a spiking study into matrix, or comparison against an orthogonal procedure. Accuracy can be inferred once precision, response within the range, and specificity are established.

Recommended data (3.3.1.4): an appropriate number of determinations and levels across the reportable range — for example 3 concentrations × 3 replicates of the full procedure. Report as mean percent recovery of a known added amount, or as the difference between the mean and the accepted true value, together with an appropriate 100(1−α)% confidence interval or justified alternative interval. The observed interval should be compatible with the accuracy criterion. For impurities, state whether the determination is weight/weight or area percent. For quantitative multivariate procedures use RMSEP, compared against an acceptable RMSEC.

Precision (3.3.2). Use authentic homogeneous samples, or artificially prepared ones if unavailable.

  • Repeatability (3.3.2.1): a minimum of 9 determinations covering the reportable range (for example 3 concentrations × 3 replicates), or a minimum of 6 determinations at 100% of the test concentration.
  • Intermediate precision (3.3.2.2): establish the effects of random events — typically different days, environmental conditions, analysts, and equipment. Studying these effects individually is not necessary, and design of experiments is encouraged. The extent should be justified from development understanding and risk assessment (ICH Q14).
  • Reproducibility (3.3.2.3): an inter-laboratory trial. Usually not required for a regulatory submission, but consider it for pharmacopoeial standardisation or multi-site procedures.

Recommended data (3.3.2.4): report the standard deviation, the relative standard deviation, and an appropriate 100(1−α)% confidence interval.

Combined accuracy and precision (3.3.3). Instead of separate criteria, assess total impact against a single combined criterion, using a prediction interval, a tolerance interval, or a confidence interval. Report the combined value, describe the approach, and supply the individual results as supplemental information where they help justify suitability.

Robustness (3.4). Deliberate variation of procedure parameters, plus stability of sample preparations and reagents over the duration of the procedure. Considered during development; may be submitted as development data case-by-case or made available on request. See ICH Q14 section 5.

Raw lines 95 to 175
176## Lifecycle, transfer, and prior knowledge
178Section 2.1 permits suitable development data (ICH Q14) to form part of the validation data, and
179allows abbreviated validation testing for an established platform procedure used for a new purpose,
180with scientific justification. A validation protocol must exist before the study, stating the
181intended purpose, the characteristics to be validated, and the associated criteria; where prior
182knowledge is used, justify it. Results are summarised in a validation report.
184The experimental design should reflect the number of replicates used in routine analysis to generate
185a reportable result, unless a different number is justified.
187Section 2.2 covers change: partial or full revalidation may be needed, decided on science and risk,
188and scoped to the characteristics the change affects. **Transfer** to another laboratory calls for
189partial or full revalidation and/or comparative analysis of representative samples; not performing
190transfer experiments requires justification. **Co-validation** across multiple sites can demonstrate
191the criteria are met and can simultaneously satisfy transfer at the participating sites.

Lifecycle, transfer, and prior knowledge

Section 2.1 permits suitable development data (ICH Q14) to form part of the validation data, and allows abbreviated validation testing for an established platform procedure used for a new purpose, with scientific justification. A validation protocol must exist before the study, stating the intended purpose, the characteristics to be validated, and the associated criteria; where prior knowledge is used, justify it. Results are summarised in a validation report.

The experimental design should reflect the number of replicates used in routine analysis to generate a reportable result, unless a different number is justified.

Section 2.2 covers change: partial or full revalidation may be needed, decided on science and risk, and scoped to the characteristics the change affects. Transfer to another laboratory calls for partial or full revalidation and/or comparative analysis of representative samples; not performing transfer experiments requires justification. Co-validation across multiple sites can demonstrate the criteria are met and can simultaneously satisfy transfer at the participating sites.

Raw lines 176 to 192
193## Annex 2 — illustrative technique examples
195Non-mandatory worked examples, useful as a starting point for the robustness parameter list:
197| Table | Technique |
198| --- | --- |
199| 3 | Quantitative separation techniques (HPLC, GC, CE) for impurities or assay, and relative-area quantitation |
200| 4 | Elemental impurities by ICP-OES or ICP-MS |
201| 5 | Dissolution with HPLC as product performance test (corrected 30 Nov 2023) |
202| 6 | Quantitative ¹H-NMR for assay of a drug substance |
203| 7 | Biological assays |
204| 8 | Quantitative PCR |
205| 9 | Particle size measurement |
206| 10 | NIR analytical procedure |
207| 11 | Quantitative LC/MS |
209From Table 3, a detail worth carrying forward: **relative response factors.** Where the analyte
210responds differently from the reference material, calculate the RRF from the appropriate ratio of
211responses under final procedure conditions and document it. **The illustrative Table 3 recommends correction when RRF is outside 0.8–1.2.** Where an impurity is overestimated, omitting the correction may be
212acceptable.

Annex 2 — illustrative technique examples

Non-mandatory worked examples, useful as a starting point for the robustness parameter list:

TableTechnique
3Quantitative separation techniques (HPLC, GC, CE) for impurities or assay, and relative-area quantitation
4Elemental impurities by ICP-OES or ICP-MS
5Dissolution with HPLC as product performance test (corrected 30 Nov 2023)
6Quantitative ¹H-NMR for assay of a drug substance
7Biological assays
8Quantitative PCR
9Particle size measurement
10NIR analytical procedure
11Quantitative LC/MS

From Table 3, a detail worth carrying forward: relative response factors. Where the analyte responds differently from the reference material, calculate the RRF from the appropriate ratio of responses under final procedure conditions and document it. The illustrative Table 3 recommends correction when RRF is outside 0.8–1.2. Where an impurity is overestimated, omitting the correction may be acceptable.

Raw lines 193 to 213
214## Multivariate procedures (2.5)
216Results come from a model relating many input variables to the property of interest. Validate in two
217phases:
2191. **Model development** — calibration plus internal testing. Test data may be a separate set or
220 part of the calibration set used rotationally, and are used to estimate performance and tune
221 parameters such as the number of PLS latent variables. See ICH Q14.
2222. **Model validation** — an independent validation set. For identification libraries, analyse
223 challenge samples *not* represented in the library to demonstrate discriminative ability.
225Samples need reference values or categories, normally from a validated or pharmacopoeial reference
226procedure whose performance **equals or exceeds** the expected performance of the multivariate
227procedure. Reference measurement and multivariate data collection should be on the same samples
228within a period short enough to assure sample and measurement stability. Describe any correlation or
229unit conversion, and any assumptions.

Multivariate procedures (2.5)

Results come from a model relating many input variables to the property of interest. Validate in two phases:

  1. Model development — calibration plus internal testing. Test data may be a separate set or part of the calibration set used rotationally, and are used to estimate performance and tune parameters such as the number of PLS latent variables. See ICH Q14.
  2. Model validation — an independent validation set. For identification libraries, analyse challenge samples not represented in the library to demonstrate discriminative ability.

Samples need reference values or categories, normally from a validated or pharmacopoeial reference procedure whose performance equals or exceeds the expected performance of the multivariate procedure. Reference measurement and multivariate data collection should be on the same samples within a period short enough to assure sample and measurement stability. Describe any correlation or unit conversion, and any assumptions.

Raw lines 214 to 229