1# ISO/IEC 17025 Process Framework (No Standard Text)
2
3Research basis: **2026-10-01**. This reference summarizes a preparation process and
4evidence architecture for testing and calibration laboratory competence work. It does
5not reproduce requirements and is not a substitute for the standard.
6

ISO/IEC 17025 Process Framework (No Standard Text)

Research basis: 2026-10-01. This reference summarizes a preparation process and evidence architecture for testing and calibration laboratory competence work. It does not reproduce requirements and is not a substitute for the standard.

Raw lines 1 to 6
7## Copyright and authorized access
8
9ISO and IEC publications are copyrighted. Obtain ISO/IEC 17025 from
10[ISO](https://www.iso.org/standard/66912.html), IEC, an ISO national member, or another
11authorized source; see [ISO copyright](https://www.iso.org/copyright.html). Do not ask
12an agent to retrieve, transcribe, summarize clause-by-clause, or store proprietary
13text. Accreditation-body checklists that quote the standard are equally copyrighted and
14are usually licensed only to the accredited laboratory.

Copyright and authorized access

ISO and IEC publications are copyrighted. Obtain ISO/IEC 17025 from ISO, IEC, an ISO national member, or another authorized source; see ISO copyright. Do not ask an agent to retrieve, transcribe, summarize clause-by-clause, or store proprietary text. Accreditation-body checklists that quote the standard are equally copyrighted and are usually licensed only to the accredited laboratory.

Raw lines 7 to 15
16## Current edition
18- [ISO/IEC 17025:2017](https://www.iso.org/standard/66912.html) is Edition 3. No
19 successor edition was identified at the research basis date. The catalogue confirms
20 the 2017 edition after its 2023 review. Its listed 2018 corrections concern French
21 and Spanish versions, not the English text.
22- The 2017 edition restructured the 2005 edition around laboratory activities and
23 risk-based thinking. Using a 2005 clause checklist as current criteria is
24 obsolete. A manual may retain an older internal organization if its controlled
25 cross-reference and implementation address the current standard; headings alone
26 do not determine conformity.
27- Record publisher, title, edition, amendments/corrigenda, authorized location, access
28 date, source owner, currency-review date, impact decision, and approval in the
29 controlled source ledger.

Current edition

  • ISO/IEC 17025:2017 is Edition 3. No successor edition was identified at the research basis date. The catalogue confirms the 2017 edition after its 2023 review. Its listed 2018 corrections concern French and Spanish versions, not the English text.
  • The 2017 edition restructured the 2005 edition around laboratory activities and risk-based thinking. Using a 2005 clause checklist as current criteria is obsolete. A manual may retain an older internal organization if its controlled cross-reference and implementation address the current standard; headings alone do not determine conformity.
  • Record publisher, title, edition, amendments/corrigenda, authorized location, access date, source owner, currency-review date, impact decision, and approval in the controlled source ledger.
Raw lines 16 to 30
31## The vocabulary trap: laboratories are accredited, not certified
33A laboratory is **accredited** to ISO/IEC 17025 by an accreditation body. It is not
34"ISO 17025 certified." A document labelled a certificate does not substitute for
35accreditation or establish accredited technical scope.
36Writing "certified" in a report, tender response, or quality manual is a substantive
37error that assessors and customers both notice.
39- Accreditation bodies operate under ISO/IEC 17011 and are peer-evaluated through the
40 recognition arrangement. Certification bodies operate under ISO/IEC 17021-1. These
41 are different schemes with different scope statements.
42- Since **2026-01-01**, Global Accreditation Cooperation Incorporated has replaced the
43 former ILAC and IAF and operates its own Multilateral Recognition Arrangement.
44 Results and certificates issued under the former ILAC MRA / IAF MLA remain
45 recognized during the transition. Verify current logo, claim, and endorsement wording
46 with the accreditation body before reproducing any of it.
47- ISO/IEC 17025 accreditation is not ISO 9001 certification, not a product approval,
48 not a regulatory authorization, and not a statement that any particular result is
49 correct.

The vocabulary trap: laboratories are accredited, not certified

A laboratory is accredited to ISO/IEC 17025 by an accreditation body. It is not "ISO 17025 certified." A document labelled a certificate does not substitute for accreditation or establish accredited technical scope. Writing "certified" in a report, tender response, or quality manual is a substantive error that assessors and customers both notice.

  • Accreditation bodies operate under ISO/IEC 17011 and are peer-evaluated through the recognition arrangement. Certification bodies operate under ISO/IEC 17021-1. These are different schemes with different scope statements.
  • Since 2026-01-01, Global Accreditation Cooperation Incorporated has replaced the former ILAC and IAF and operates its own Multilateral Recognition Arrangement. Results and certificates issued under the former ILAC MRA / IAF MLA remain recognized during the transition. Verify current logo, claim, and endorsement wording with the accreditation body before reproducing any of it.
  • ISO/IEC 17025 accreditation is not ISO 9001 certification, not a product approval, not a regulatory authorization, and not a statement that any particular result is correct.
Raw lines 31 to 50
51## Hard boundary
53This skill and its files cannot:
55- accredit a laboratory, issue or validate an accreditation certificate or schedule,
56 or predict an assessment outcome;
57- decide which accreditation scheme, jurisdiction requirement, or customer
58 specification applies;
59- replace the laboratory's technical management, quality manager, authorized signatory,
60 accreditation body, assessor, or proficiency-testing provider;
61- validate a method, compute or approve a measurement uncertainty budget, establish
62 metrological traceability, or judge whether a decision rule is fit for purpose; or
63- infer competence from a document title, keyword, template, checklist, or script
64 result.
66Use outputs as a list of evidence questions for accountable human review.

Hard boundary

This skill and its files cannot:

  • accredit a laboratory, issue or validate an accreditation certificate or schedule, or predict an assessment outcome;
  • decide which accreditation scheme, jurisdiction requirement, or customer specification applies;
  • replace the laboratory's technical management, quality manager, authorized signatory, accreditation body, assessor, or proficiency-testing provider;
  • validate a method, compute or approve a measurement uncertainty budget, establish metrological traceability, or judge whether a decision rule is fit for purpose; or
  • infer competence from a document title, keyword, template, checklist, or script result.

Use outputs as a list of evidence questions for accountable human review.

Raw lines 51 to 67
68## Scope of accreditation is per activity, not per organization
70The accredited scope is defined item by item: field or discipline, method or procedure
71and its issue, the measurand or property determined, and the range with the reported
72uncertainty basis. For calibration laboratories this is normally expressed through
73calibration and measurement capability statements.
75Consequences for evidence preparation:
77- Work outside the accredited scope must not carry an accreditation claim or
78 endorsement, even at an accredited site under an accredited quality system.
79- A method change, a range extension, a new location, or a new authorized signatory is
80 a scope question, not only an internal change-control question.
81- Subcontracted work has its own rules for customer notification and for the
82 subcontractor's own accreditation status. Record both.
84Capture scope items with:
86```bash
87PYTHONDONTWRITEBYTECODE=1 python3 scripts/validate_scope_intake.py \
88 /path/to/laboratory-scope-intake.json --standard iso-17025
89```
91Copy `assets/templates/laboratory-scope-intake-template.json` outside the skill first.
92The distributed template fails closed by design.

Scope of accreditation is per activity, not per organization

The accredited scope is defined item by item: field or discipline, method or procedure and its issue, the measurand or property determined, and the range with the reported uncertainty basis. For calibration laboratories this is normally expressed through calibration and measurement capability statements.

Consequences for evidence preparation:

  • Work outside the accredited scope must not carry an accreditation claim or endorsement, even at an accredited site under an accredited quality system.
  • A method change, a range extension, a new location, or a new authorized signatory is a scope question, not only an internal change-control question.
  • Subcontracted work has its own rules for customer notification and for the subcontractor's own accreditation status. Record both.

Capture scope items with:

PYTHONDONTWRITEBYTECODE=1 python3 scripts/validate_scope_intake.py \
  /path/to/laboratory-scope-intake.json --standard iso-17025

Copy assets/templates/laboratory-scope-intake-template.json outside the skill first. The distributed template fails closed by design.

Raw lines 68 to 93
Runs code python3 scripts/validate_scope_intake.py :87
94## Process and evidence domains
96The `iso-17025` profile carries these domain labels for manifests and gap reports.
97They are workflow topics, not clause references:
99`scope-and-impartiality`, `organizational-structure-and-management`,
100`personnel-competence`, `facilities-and-environmental-conditions`,
101`equipment-and-calibration`, `metrological-traceability`,
102`externally-provided-products-and-services`, `review-of-requests-and-contracts`,
103`method-selection-verification-and-validation`, `sampling`, `handling-of-items`,
104`technical-records`, `measurement-uncertainty`,
105`validity-of-results-and-proficiency-testing`, `reporting-and-decision-rules`,
106`complaints`, `nonconforming-work`, `data-and-information-management`,
107`internal-audit`, `management-review`, `corrective-action-and-improvement`.
109Each domain needs an owner, status, evidence IDs, source/version reference, recorded
110approval, and links to open gaps. Sample **records**, not only procedures: a procedure
111describing uncertainty evaluation is not evidence that a budget exists for a reported
112measurand.

Process and evidence domains

The iso-17025 profile carries these domain labels for manifests and gap reports. They are workflow topics, not clause references:

scope-and-impartiality, organizational-structure-and-management, personnel-competence, facilities-and-environmental-conditions, equipment-and-calibration, metrological-traceability, externally-provided-products-and-services, review-of-requests-and-contracts, method-selection-verification-and-validation, sampling, handling-of-items, technical-records, measurement-uncertainty, validity-of-results-and-proficiency-testing, reporting-and-decision-rules, complaints, nonconforming-work, data-and-information-management, internal-audit, management-review, corrective-action-and-improvement.

Each domain needs an owner, status, evidence IDs, source/version reference, recorded approval, and links to open gaps. Sample records, not only procedures: a procedure describing uncertainty evaluation is not evidence that a budget exists for a reported measurand.

Raw lines 94 to 113
114## The four technical areas that carry most findings

The four technical areas that carry most findings

Raw lines 114 to 115
116### 1. Metrological traceability
118Traceability is a documented, unbroken chain to a stated reference with stated
119uncertainty at each step — not a drawer of calibration certificates. The governing
120policy is **ILAC P10:07/2020** (implementation July 2021, revised for the 2017
121edition), still carried in the current Global ACI library. Its proposed successor
122Global ACI-TECH-1-011 remains in process in the September 2026 cross-reference
123table; do not cite that identifier as a published policy.
125Evidence to assemble per measurement: the stated reference, the calibration provider
126and its accreditation status and scope, certificate identity and date, the uncertainty
127contribution carried forward, interval justification, and intermediate-check records.

1. Metrological traceability

Traceability is a documented, unbroken chain to a stated reference with stated uncertainty at each step — not a drawer of calibration certificates. The governing policy is ILAC P10:07/2020 (implementation July 2021, revised for the 2017 edition), still carried in the current Global ACI library. Its proposed successor Global ACI-TECH-1-011 remains in process in the September 2026 cross-reference table; do not cite that identifier as a published policy.

Evidence to assemble per measurement: the stated reference, the calibration provider and its accreditation status and scope, certificate identity and date, the uncertainty contribution carried forward, interval justification, and intermediate-check records.

Raw lines 116 to 128
129### 2. Measurement uncertainty
131For calibration, use Global ACI-TECH-1-009 (M), version 1.0 (2026-06-26),
132successor to ILAC P14, for the policy basis. A scope CMC and a delivered
133calibration's uncertainty are distinct. Record which measurands have evaluated
134uncertainty, the evaluation basis, the
135contributions included, how uncertainty is reported, and how it interacts with the
136decision rule. An uncertainty budget that exists only in a spreadsheet with no
137controlled revision, owner, or approval is a gap, not evidence.
139This skill does not compute uncertainty. For the repo's uncertainty tooling, see the
140`uncertainty-and-units` skill; its outputs still require authorized technical review
141before entering a controlled record.

2. Measurement uncertainty

For calibration, use Global ACI-TECH-1-009 (M), version 1.0 (2026-06-26), successor to ILAC P14, for the policy basis. A scope CMC and a delivered calibration's uncertainty are distinct. Record which measurands have evaluated uncertainty, the evaluation basis, the contributions included, how uncertainty is reported, and how it interacts with the decision rule. An uncertainty budget that exists only in a spreadsheet with no controlled revision, owner, or approval is a gap, not evidence.

This skill does not compute uncertainty. For the repo's uncertainty tooling, see the uncertainty-and-units skill; its outputs still require authorized technical review before entering a controlled record.

Raw lines 129 to 142
Names a skill uncertainty-and-units :140
143### 3. Method selection, verification, and validation
145For a standard method used as published, record verification that the laboratory
146can achieve the required performance before use; an unmodified method is not an
147exemption from verification. Distinguish modified, out-of-scope, non-standard, and
148laboratory-developed methods, and document the validation needed for their intended
149use. The evidence expected differs in each case.
150Record the performance characteristics evaluated, the acceptance criteria, the data,
151and the authorized approval to put the method into service.

3. Method selection, verification, and validation

For a standard method used as published, record verification that the laboratory can achieve the required performance before use; an unmodified method is not an exemption from verification. Distinguish modified, out-of-scope, non-standard, and laboratory-developed methods, and document the validation needed for their intended use. The evidence expected differs in each case. Record the performance characteristics evaluated, the acceptance criteria, the data, and the authorized approval to put the method into service.

Raw lines 143 to 152
153### 4. Decision rules and statements of conformity
155If reports state conformity to a specification, the decision rule must be documented,
156agreed with the customer where required, and applied consistently — including how
157measurement uncertainty is treated at the specification limit. The guidance basis is
158**Global ACI-TECH-1-002 (G), version 1.0** (2026-06-26), the successor to
159ILAC G8:09/2019. JCGM 106:2012 / ISO/IEC Guide 98-4:2012 is the metrological
160companion. Confirm the scheme's adopted basis and current issues before controlled use.
162Assemble, per scope item that reports conformity: the rule, the agreement record, the
163uncertainty treatment, the report wording, and the authorized signatory.

4. Decision rules and statements of conformity

If reports state conformity to a specification, the decision rule must be documented, agreed with the customer where required, and applied consistently — including how measurement uncertainty is treated at the specification limit. The guidance basis is Global ACI-TECH-1-002 (G), version 1.0 (2026-06-26), the successor to ILAC G8:09/2019. JCGM 106:2012 / ISO/IEC Guide 98-4:2012 is the metrological companion. Confirm the scheme's adopted basis and current issues before controlled use.

Assemble, per scope item that reports conformity: the rule, the agreement record, the uncertainty treatment, the report wording, and the authorized signatory.

Raw lines 153 to 164
165## Validity of results and proficiency testing
167Interlaboratory comparison and proficiency-testing participation is monitoring
168evidence, and a passing result is not a competence conclusion. Record the plan (which
169scope items, what frequency, what provider), each result, the evaluation against
170acceptance criteria, and — most importantly — the investigation and corrective action
171for any questionable or unsatisfactory outcome. An unsatisfactory result with no
172documented investigation is a blocker; treat it as one.
174Where the laboratory cannot participate because no scheme exists for a measurand,
175record the alternative approach and its authorization rather than leaving a silence.

Validity of results and proficiency testing

Interlaboratory comparison and proficiency-testing participation is monitoring evidence, and a passing result is not a competence conclusion. Record the plan (which scope items, what frequency, what provider), each result, the evaluation against acceptance criteria, and — most importantly — the investigation and corrective action for any questionable or unsatisfactory outcome. An unsatisfactory result with no documented investigation is a blocker; treat it as one.

Where the laboratory cannot participate because no scheme exists for a measurand, record the alternative approach and its authorization rather than leaving a silence.

Raw lines 165 to 176
177## Shared checks that apply here
179- `scripts/audit_document_records.py` — controlled documents, technical records,
180 retention basis, and external-source currency. Standard-agnostic.
181- `scripts/check_capa.py` — nonconforming work and corrective action, including
182 effectiveness evidence before closure.
183- `scripts/check_supplier_controls.py` — externally provided products and services,
184 including calibration providers, reference-material suppliers, and subcontracted
185 laboratories.
186- `scripts/validate_evidence_manifest.py` and `scripts/gap_analyzer.py` with
187 `--standard iso-17025` — bounded evidence manifest and domain-level gap view.
189`scripts/check_traceability.py` and `scripts/check_qmsr_transition.py` are
190device-lifecycle checks and do not apply to laboratory accreditation work. Note that
191`check_traceability.py` concerns design/risk traceability, **not** metrological
192traceability — the words collide and the check is the wrong tool here.

Shared checks that apply here

  • scripts/audit_document_records.py — controlled documents, technical records, retention basis, and external-source currency. Standard-agnostic.
  • scripts/check_capa.py — nonconforming work and corrective action, including effectiveness evidence before closure.
  • scripts/check_supplier_controls.py — externally provided products and services, including calibration providers, reference-material suppliers, and subcontracted laboratories.
  • scripts/validate_evidence_manifest.py and scripts/gap_analyzer.py with --standard iso-17025 — bounded evidence manifest and domain-level gap view.

scripts/check_traceability.py and scripts/check_qmsr_transition.py are device-lifecycle checks and do not apply to laboratory accreditation work. Note that check_traceability.py concerns design/risk traceability, not metrological traceability — the words collide and the check is the wrong tool here.

Raw lines 177 to 193
Runs code scripts/audit_document_records.py :179
Runs code scripts/check_capa.py :181
Runs code scripts/check_supplier_controls.py :183
Runs code scripts/validate_evidence_manifest.py :186
Runs code scripts/gap_analyzer.py :186
Runs code scripts/check_traceability.py :189
Runs code scripts/check_qmsr_transition.py :189
194## Common preparation failures
196- Claiming "ISO 17025 certified" instead of accredited, or attaching an accreditation
197 symbol to results outside the accredited scope.
198- Treating calibration certificates as traceability evidence without the uncertainty
199 chain, provider scope, or interval justification.
200- Reporting conformity with no documented decision rule, or a rule that ignores
201 uncertainty at the limit.
202- Verifying a standard method once at introduction and never revisiting it after a
203 method issue, instrument, or personnel change.
204- Recording proficiency-testing scores while leaving unsatisfactory results
205 uninvestigated.
206- Presenting an impartiality statement with no identified risks, no controls, and no
207 review.
208- Filing a document register as the deliverable when no technical records were sampled.

Common preparation failures

  • Claiming "ISO 17025 certified" instead of accredited, or attaching an accreditation symbol to results outside the accredited scope.
  • Treating calibration certificates as traceability evidence without the uncertainty chain, provider scope, or interval justification.
  • Reporting conformity with no documented decision rule, or a rule that ignores uncertainty at the limit.
  • Verifying a standard method once at introduction and never revisiting it after a method issue, instrument, or personnel change.
  • Recording proficiency-testing scores while leaving unsatisfactory results uninvestigated.
  • Presenting an impartiality statement with no identified risks, no controls, and no review.
  • Filing a document register as the deliverable when no technical records were sampled.
Raw lines 194 to 209
210## Sources
212- [Global ACI document library](https://global-aci.org/en/global_aci-documents/) — current policy/guidance designations
213- [BIPM JCGM publications](https://www.bipm.org/en/committees/jc/jcgm/publications) — JCGM 106:2012
214- [ISO/IEC 17025:2017](https://www.iso.org/standard/66912.html) — catalogue entry
215- [ISO/IEC 17025 landing page](https://www.iso.org/ISO-IEC-17025-testing-and-calibration-laboratories.html)
216- [ILAC policy series](https://ilac.org/publications-and-resources/ilac-policy-series/) — legacy policy archive; P10 is still carried by Global ACI
217- [GLOBAC launch](https://iaf.nu/en/news/global-accreditation-cooperation-incorporated-launch-unifies-international-accreditation-organisations-and-strengthens-worldwide-trust/)
218- [Specifying use of GLOBAC accreditation](https://ilac.org/latest_ilac_news/iaf-and-ilac-release-information-on-specifying-use-of-globac-accreditation/)
219- `references/source-ledger.md` — dated baseline and provenance limitations
220- `references/assurance-lanes.md` — how this lane differs from certification and
221 regulatory inspection

Sources

Raw lines 210 to 221